6.0 
CAM Pressure Remote Installation and Mounting 
 
Please apply precautionary safety procedures during installation.  The  CAM  Pressure  Remote  has  been 
specifically designed to meet IP-65 standard (protected from low pressure water jets from any direction, 
limited ingress protection (IP)). The CAM Pressure Remote can be installed in locations where it will be 
intermittently exposed to spraying water, rain or high humidity.  
 
 
CAUTION: The CAM Pressure Remote should never be submerged or used under water. 
 
6.1 
Operating Temperature 
 
The CAM Pressure Remote should only be installed and operated in locations that maintain an ambient 
temperature between -20°C to 70°C (-4°F to 158°F). 
 
6.2 
CAM Pressure Remote Installation with a Helium Regulator 
 
Please apply precautionary safety procedures during installation. This involves the following steps: 
•  Attach helium regulator to cylinder. Note: Do not carry the regulator by the CAM  Pressure 
Remote. This could cause undue stress to the unit 
•  Only use the hand tight wheel on helium regulator to tighten to the cylinder 
•  Once the hand tight wheel is tight apply pressure from the cylinder 
 
For future use, make note of the 16 digit identification code on the label of the CAM Pressure Remote.  
 
Note: Visually check O ring on stem of regulator for wear with each cylinder change to prevent helium 
loss (The O ring is usually the only failed part of regulator – replace as necessary).  
 
6.3 
CAM Pressure Remote Installation with an Industrial Regulator 
 
The  CAM  Pressure  Remote  comes standard with either a 1/4 or 1/8 NPT connection. Good piping 
practices are required. Always use Teflon tape or pipe sealant on the gauge threads. It is recommended to 
use a wrench on the wrench flat of the gauge to tighten the gauge to the process. 
 
 
CAUTION: Never tighten gauge threads by holding the body of the CAM Pressure Remote
Doing so will damage the gauge and make the gauge inoperable.