3.
Device does not communicate with the computer. Remove and re-insert the USB of the
Wireless Receiver
. If this does ’t ork, restart the PC. If this does ’t ork, repla e the U“B
receiver. If this does not work replace the device.
4.
Device does not communicate with the computer. Check for and turn off all other wireless
devices in the vicinity. If the XO
1
Knee Balancing System begins to communicate, turn on the
other wireless devices sequentially to identify the source of interference. Either keep the
interfering device turned off, or move it away from the vicinity of the XO
interfering device turned off, or move it away from the vicinity of the XO
1
Knee Balancing
System
5.
Controller and USB receiver LEDs light up but PC does not display the data. Check that the serial
number on the controller label matches the serial number selected on the PC display and that
the appropriate COM Port is selected. If the serial numbers match, restart the PC. If the PC
does not display data after restarting, replace the device with a new one.
the appropriate COM Port is selected. If the serial numbers match, restart the PC. If the PC
does not display data after restarting, replace the device with a new one.
Device Accessories (detachable parts):
Leg Strap
Control Syringe, 20 ml
Computer with GUI Software
Wireless Tool, Programmed
Detachable Parts:
Vented Male Luer Lock Cap
Battery Cap Assembly
Dual Check Valve
Bonded Pressure Monitoring Extension Line
Device Specifications:
IP rating for the device: IPX2
Applied Parts: Type B
FCC ID: 2ALBY-XO1
Storage Conditions:
Storage Conditions:
Products must be stored in a clean, dry environment.
Temperature Limits: -20°C to 60°C
Manufactured for XpandOrtho, Inc.
2223 Avenida de la Playa, Ste 203, La Jolla, CA 92037-3218
2223 Avenida de la Playa, Ste 203, La Jolla, CA 92037-3218