XpandOrtho Inc XO1 사용자 설명서

다운로드
페이지 13
13 of 11 
 
 
3.
  Device does not communicate with the computer.  Remove and re-insert the USB of the 
Wireless Receiver
.  If this does ’t  ork, restart the PC.  If this does ’t  ork, repla e the U“B 
receiver.   If this does not work replace the device.    
4.
  Device does not communicate with the computer.  Check for and turn off all other wireless 
devices in the vicinity.  If the XO
1
 Knee Balancing System begins to communicate, turn on the 
other wireless devices sequentially to identify the source of interference.  Either keep the 
interfering device turned off, or move it away from the vicinity of the XO
1
 Knee Balancing 
System 
5.
  Controller and USB receiver LEDs light up but PC does not display the data.  Check that the serial 
number on the controller label matches the serial number selected on the PC display and that 
the appropriate COM Port is selected.  If the serial numbers match, restart the PC.  If the PC 
does not display data after restarting, replace the device with a new one.   
 
Device Accessories (detachable parts): 
Leg Strap   
Control Syringe, 20 ml 
Computer with GUI Software 
Wireless Tool, Programmed 
 
Detachable Parts: 
Vented Male Luer Lock Cap 
Battery Cap Assembly 
Dual Check Valve 
Bonded Pressure Monitoring Extension Line 
 
Device Specifications: 
IP rating for the device:  IPX2 
Applied Parts:  Type B  
FCC ID:  2ALBY-XO1  
 
Storage Conditions: 
Products must be stored in a clean, dry environment.    
Temperature Limits: -20°C to 60°C    
Manufactured for XpandOrtho, Inc.                                                                                             