Charm Engineering Co. Ltd CAI001-BT33 사용자 설명서

다운로드
페이지 66
 
CORITICAL ACTIVITY INDEX 
|
 53
 
 
 
3.9  Termination of CAI 
 
When intending to terminate use of CAI monitoring system, 
• 
Push the CAIs sensor connected to the CAIx connector with hand and pull it to 
disconnect. In order to avoid problems in power supply and signal processing for next 
use, do not pull the cable to disconnect it. 
• 
Remove CAIs sensors from the patient. 
• 
CAIs is not reusable as it is produced for single use. 
• 
Store CAIx in state that it is connected with CAIx cable to prevent loss of components. 
• 
When use of CAI (surgery) is continuous, maintain stand-by mode on CAI monitor, and 
connect new CAIs for making it immediately usable state.  
• 
When it is the last surgery of that day, disconnect CAIs sensor from CAIx connector and 
store the connecting cable and CAIx. 
• 
 After surgery, arrange the CAI system and store it in a designating place for next 
surgery. (When the device is contacted and/or contaminated with blood or other drugs, 
clean them up prior to storage and store the CAI in a designating place) 
 
※ Caution: Do not disconnect power cord with wet hand. Check if the hands are clean and dry 
before touching the power cord.