Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
47 
 
during active 
alarm 
Monitored parameters and alarms that use the flow measurement such as tidal volume will not 
function. 
Monitored parameters and alarms that use the pressure measurement will continue to function. 
 
E
XTERNAL 
F
LOW 
S
ENSOR 
R
EVERSED
 
Priority 
High 
Why it occurs 
The external flow sensor is connected backwards during active therapy. 
What to do 
Check the sensor position. The arrow direction should align with the air delivered to the patient. 
Requirements 
Active flow or dual limb circuit 
Device 
performance 
during active 
alarm 
In volume modes, device will not accurately control volume. The peak inspiratory pressure will 
continue to increase. 
In pressure modes, therapy will be unaffected. 
All monitored parameters and alarms will continue to function. 
 
A
CTIVE 
E
XHALATION 
V
ALVE 
F
AILED
 
Priority 
High 
Why it occurs 
The active exhalation valve is stuck closed. 
What to do 
• 
For a single-limb active circuit, check all connections to the valve and check that the valve is clear. 
• 
For a dual limb circuit, check that the valve is clear.. 
Requirements 
Active flow or dual limb circuit 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved and the device continues to operate. 
 
 
C
HECK 
A
CTIVE 
E
XHALATION 
P
RESSURE 
C
ONTROL 
L
INE
 
Priority 
High 
Why it occurs 
The active-exhalation pressure control line is not connected, becomes disconnected, or contains water 
droplets that affect the active exhalation valve line pressure reading. 
What to do 
Inspect the line. Empty or replace it if necessary. 
Requirements 
Active flow or dual limb circuit 
Device 
performance 
during active 
alarm 
Therapy delivery will be compromised.   
All monitored parameters and alarms will continue to function. 
 
P
ROXIMAL 
P
RESSURE 
L
INE 
D
ISCONNECTED
 
Priority 
High 
Why it occurs 
The proximal pressure line is not connected. 
What to do 
• 
Verify the proximal pressure line is connected properly at both ends and the line is clean and 
untangled.  
• 
Ensure the main circuit is connected properly and does not have large leaks.  
• 
Ensure the exhalation valve is intact. 
Requirements 
Dual limb, active PAP, or active flow circuit 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the proximal pressure line is connected properly. 
 
The device continues to operate. 
 
O
XYGEN 
R
EGULATION
 
Priority 
High 
Why it occurs 
The flow measured FiO
2
 is not within 10% ± 3% of the FiO
2
 setting. 
What to do 
• 
Verify the oxygen blender is connected properly.