Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
56 
 
L
OW 
F
I
O
2
 
Priority 
Medium 
Why it occurs 
The measured FiO
2
 is less than or equal to the low FiO
2
 alarm setting for 10 seconds in active therapy. 
What to do 
• 
Confirm proper connection of the oxygen source.  
• 
When using low-flow oxygen, check for unnecessary leaks. Increase flow rate if possible.  
• 
When using high pressure oxygen, check for excessive leaks. 
• 
If the alarm persists, calibrate the oxygen sensor. See “chapter 7, Device Options.” 
Requirements 
FiO
2
 sensor 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured FiO
2
 exceeds the low FiO
2
 alarm setting in 
active therapy. 
 
Device will continue to operate. 
Alarm Settings 
21 to 95% in increments of 1%. 
 
H
IGH 
F
I
O
2
 
Priority 
Medium 
Why it occurs 
The measured FiO
2
 is greater than or equal to the high FiO
2
 alarm setting for 10 seconds. 
What to do 
• 
Confirm proper connection of the oxygen source. 
• 
When using low-flow oxygen, decrease the flow rate if possible.  
• 
If the alarm persists, calibrate the oxygen sensor. See “chapter 7, Device Options.” 
Requirements 
FiO
2
 sensor 
100% O
2
 must not be active. This alarm is disabled when 100% O
2
 is active 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured FiO
2
 falls below the high FiO
2
 alarm setting in 
active therapy (for 10 consecutive seconds). 
 
Device will continue to operate. 
Alarm Settings 
27 to 100% in increments of 1% 
 
Low-priority System Alarms 
S
TUCK 
K
EY
 
Priority 
Medium 
Why it occurs 
A stuck On/Off (Standby) button or Alarm Silence button has been detected for at least 120 seconds. 
What to do 
Contact Philips Respironics Customer Service 
Device 
performance 
during active 
alarm 
Device will continue to operate. 
 
 
I
NLET 
F
ILTER 
B
LOCKED
 
Priority 
Low 
Why it occurs 
The inlet filter becomes blocked and delivered therapy is reduced. 
What to do 
Remove the filter. If the filter is the air-inlet foam filter, rinse it according to the instructions in the 
“Cleaning and Disinfection” chapter. Otherwise, replace the filter. 
Requirements 
Inlet filter installed 
Device 
performance 
during active 
alarm 
Device will continue to operate. 
 
Algorithm 
summary  
Because the inlet filter is blocked, the pressure generated by the ventilator is less than 75% of the 
estimated outlet pressure based on the nominal performance characteristics of the device.  
 
C
HECK 
P
ROXIMAL 
P
RESSURE 
L
INE
 
Priority 
Low