Instructions for Use | Alarms and System Messages 
 
 
 
Testing Alarms 
60 
 
Prescription Alarm Availability by Therapy Mode 
Most alarms operate regardless of the therapy mode being delivered. However, certain alarms operate only when certain therapy modes are being 
delivered.  
The chart below lists those alarms and the therapy modes in which the alarms are available. 
Alarm 
A/C-PC 
A/C-VC 
CPAP 
PSV 
S/T 
SIMV-PC 
SIMV-VC 
Apnea 
(requires Backup Ventilation) 
Circuit Disconnect 
High Tidal Volume 
Low Tidal Volume 
High Minute Ventilation 
Low Minute Ventilation 
High Respiratory Rate 
Low Respiratory Rate 
High Inspiratory Pressure 
 
 
 
 
 
Low Inspiratory Pressure 
 
 
 
 
 
High SpO
2
 
Low SpO
2
 
High Pulse Rate 
Low Pulse Rate 
High etCO
2
 
Low etCO
2
 
High FiO
2
 
Low FiO
2
 
Testing Alarms 
Test alarms any time you make a significant change to the system.  
Testing Circuit Disconnection Alarms 
For ventilator-dependent patients, do not rely on any single alarm to detect when a circuit is disconnected. One or more of the following alarms 
may indicate a disconnected circuit.  
Circuit Disconnected  
Low Tidal Volume 
Low Minute Ventilation 
Low Respiratory Rate 
Low Peak Inspiratory Pressure alarms (user settable for volume modes) 
Leakage alarm (Active PAP circuit only) 
 
To test that these alarms detect a circuit disconnection: 
1.  Ensure that the patient is connected to the ventilator and that ventilation is stabilized. Ensure that none of the alarms above are active. 
2.  Disconnect the circuit at the patient end of the circuit. Remove the leak device (passive circuits) and any humidifiers or other circuit 
accessories at the patient end of the circuit. Only the circuit tube should remain. 
3.  Confirm that one or more of the above-listed alarms activate.  
4.  Reconnect the circuit and confirm that any active alarm automatically resets. 
Testing Circuit Obstruction Alarms 
For ventilator-dependent patients, do not rely on any single alarm to detect when a circuit is obstructed. One or more of the following alarms may 
indicate an obstructed circuit.  
Obstruction 
High Inspiratory Pressure