XRpad2 4343
For Your Safety
WARNING
The detector is not designed to control X-ray dose. The system 
integrator is responsible for controlling the X-ray radiation.
WARNING
Connection of the X-ray detector to the installed IT network that 
includes other equipment could result in previously unidentified risk 
to patients, operators or third parties. the responsible organization 
should identify, analyze, evaluate and control these risk. Subsequent 
changes to the IT-network could introduce also new risks and require 
additional analysis.
Caution
Do not operate the X-ray detector in a location with the following 
conditions.
Close to fluid or places where fluid is used
Close to heat sources, such as a heater
High temperature environment
High humidity environment
High condensation environment
Extreme cold environment
Dusty environment
Salty or sulphurous environment
Near a vibrating environment
Environment where there is insufficient air circulation to enable the X-ray 
detector, and power supply to dissipate heat.
Ignoring this caution may result in personal injury or damage to 
the product.
Caution
When using the X-ray detector in Automatic Exposure Detection 
(AED) mode, the temperature may increase due to the heat generated 
by the X-ray detector electronics to support the AED mode. Follow the 
integration guide and instructions for the use of the X-ray detector in 
AED mode. Ignoring this caution may result in personal injury or 
damage to the product.
Caution
Monitor the temperature of the X-ray detector and allow the detector to 
cool down when the temperature of the detector gets hot. Ignoring this 
caution may result in personal injury or damage to the product.
Caution
An Image Quality Test or Customer Acceptance test should be 
performed before the detector is used for the first time with patients. 
Version 
01 
2017-11-03