Centauri Medical Inc. 1116AG 사용자 설명서
Centauri Medical, Inc.
Document No. 1312AJ
DynaSense Instructions For Use
Page 52 of 54
TECHNICAL SPECIFICATIONS
Electrical Voltage (USB RF Transceiver)
5 VDC
Electrical Voltage (Relay Antenna)
120 VAC
Electrical Voltage (Relay Antenna Power
Adapter)
100-240V ~ 50/60Hz, .18A,
Output: 5.0V DC 1.0 A
Electrical Voltage (Patient Sensor)
3 VDC
Patient Sensor Power Source
Internally Powered
Patient Sensor Size
1.8” x 2.0”,
adhesive extends to 2.4” x 2.8”
Patient Sensor Weight (max)
20 g
Patient Sensor Skin Adhesive Material
Acrylic adhesive on
polyurethane carrier
Patient Sensor Applied Part Type
Type BF Applied Part
Patient Sensor Cover Material
Polystyrene
Optimal Distance between Patient Sensor
and Relay Antenna
15 feet
Wireless Transmission Protocol
IEEE 802.15.4 Compliant
Wireless Transmission Frequency
2.4 GHz
Relay Antenna Classification
Class II Medical Electrical Equipment,
IEC 60601
Ingress Protection (IP) Rating
IPX7 for Patient Sensor, IP2X for Relay
Antenna and USB RF Transceiver
Intended System Operation
Continuous Operation