Setup
4. Setup
 WARNING: These requirements must be met when Sonicaid 
Freedom is connected to any other electrical equipment.
1  
Medical equipment must comply with IEC60601-1, or   
 
  
equivalent.
2   
The confi gured system must comply with the requirements of     
   
IEC60601-1:2005; clause 16.
4.1 Receiver 
Connection
The receiver unit should be plugged into the same mains 
supply circuit as other equipment in use on the same patient. 
Do not make any modifi cations to the power supply of the 
receiver and transducer units.
• 
Connect the interface cable which is clearly identifi ed, from output of the 
receiver unit to the interface input of the fetal monitor.
• 
Connect the receiving antenna and select RF channel (See Section 8.3) 
• 
Dock the transducers with the receiver powered off.
• 
Connect the mains supply lead from the receiver unit to mains supply. This 
will pair the receiver and transducers.
• 
Charge the transducer batteries
• 
Switch on the FM800E monitor (refer to its Instructions for Use)
There is no end user tuning required other than changing the 
channel. The user selects a channel on the receiver rotary 
switches. The user docks an ultrasound transducer to the 
receiver within 20 seconds of a receiver power up to pair it to 
the same channel as the receiver.
The system is ready for monitoring.