Huntleigh Healthcare Limited Huntleigh Diagnostic Products Division SF1US 사용자 설명서

다운로드
페이지 48
20
Oper
ation
5. Operation 
5.1 Getting 
Started
• 
Charge the transducer batteries
• 
Switch on the FM800E monitor (refer to its Instructions for Use)
• 
Apply transducers to the patient (Refer to Section 5.2).
When the US-transducer is undocked from the charging bay, the unit is 
automatically switched on and is ready to use. 
The system is ready for monitoring if the transmission indicator LED and battery 
indicator LEDs are illuminated green.
5.2  Application of the Transducers
Ultrasound transducer
Fix the transducer to the belt clip.  Apply ultrasound gel to the surface of the 
transducer.
Note: Use recommended gel only. Do not use oil based gels.
Use the minimum amount of gel required, to prevent it from sliding too easily on 
the skin.
• 
When the transducer is to be used in water, use only a little gel, or no gel if 
possible.
• 
Position the transducer. Adjust for best, clear, fetal heart sounds.  Refer to 
the FM800E instructions for use for further information. 
• Secure 
fi rmly with the belt.  For optimum performance during periods of 
mobility, ensure that the belt is tight enough to prevent the transducer 
slipping. Avoid patient discomfort through excessive tightening.