6.  Care and Cleaning
6.1 General 
Care
Although the Freedom is robust and has been designed to withstand normal 
clinical use, the unit contains delicate components, which should be handled 
and treated with care.
Periodically, and whenever the integrity of the system is in doubt, carry out a 
check of all functions as described in Section 4.2.  If there are any defects to 
the housing of a transducer do not use it in water. Contact Huntleigh or your 
distributor to get it repaired or to order a replacement.
If a transducer is dropped, check that the seal and the housing are not 
damaged. If you are in doubt please contact Huntleigh or your distributor for 
further instructions.
Please ensure that you check with your facility’s local infection 
control policy and medical equipment cleaning procedures.
Observe warnings and guidance on cleaning fl uid labelling 
regarding use and personal protective equipment (PPE).
Do not use abrasive cloths or cleaners on the transducer, 
receiver or accessories.
Do not use automatic washers or autoclaves to clean the 
transducers.
Phenolic detergent based disinfectants, solutions containing 
cationic surfactants, ammonia based compounds or perfumes 
and antiseptic solutions such as Steriscol or Hibiscrub should 
never be used on any part of the Freedom system.
If detergent or disinfectant wipes are used to clean the 
transducers or the receiver ensure that excess solution is 
squeezed from the wipe prior to use.
Always switch off the Receiver by disconnecting the unit from 
the AC supply before cleaning and disinfecting, do not allow 
any fl uid to enter the Receiver and do not immerse the receiver 
in any solution.