ouble Shooting
8. Trouble 
Shooting
This section gives some of the more common problems encountered during use 
together with possible causes.  If the problem cannot be located after consulting 
the table below, the receiver should be disconnected from the mains power 
source and a qualifi ed technician should be consulted.
Before attempting trouble-shooting, verify that all cables are properly connected 
to both the receiver, FM800E and the main power source.
Transmission Indicator LED always amber, even though the US-transducer is 
removed.
POSSIBLE CAUSE
SOLUTION
Transducer battery discharged.
Charge transducer
Receiver and transducer are 
working on different RF-
channels.
Select the correct transducer
Re-pair the transducers to receiver. (See 
section 4).
Transmission Indicators fl icker sporadically amber/green when patient is walking.
POSSIBLE CAUSE
SOLUTION
Patient out of range.
Ensure the patient understands the range 
limits and any local "dead spots" to avoid.
No signal on Fetal Monitor even though the indicators on receiver are green.
POSSIBLE CAUSE
SOLUTION
Interface cable to Fetal Monitor 
is disconnected or broken.
Connect cable or repair.
Interrupted recording of fetal heart rate or uterine activity.
POSSIBLE CAUSE
SOLUTION
Transducer position.
Check position of transducers for best signal.
Transducer is sliding on the 
skin.  
Reposition and fasten securely. Use less gel 
on the US transducer.
Excessive movement
Ask the patient to restrict movement.
RF-interference or patient at 
the end of range.
Instruct patient to stay inside the area where 
reception is good. If necessary change RF 
channel. Refer to 8.3.
Charging indicator is off even though the transducer is docked for charging.
POSSIBLE CAUSE
SOLUTION
Transducer not docked properly.
Re-position the transducer.
Receiver not connected to the 
mains power supply.
Connect the receiver.