10. Electromagnetic Compatibility
Make sure the environment in which the Sonicaid™ Freedom is installed 
is not subject to strong sources of electromagnetic interference (e.g. radio 
transmitters, mobile phones).
This equipment generates and uses radio frequency energy. If not installed 
and used properly, in strict accordance with the manufacturer’s instructions, 
it may cause or be subject to interference. Type-tested in a fully confi gured 
system, it complies with EN60601-1-2, the standard intended to provide 
reasonable protection against such interference. Whether the equipment causes 
interference may be determined by turning the equipment off and on. If it does 
cause or is affected by interference, one or more of the following measures may 
correct the interference:
  • 
Re-orienting the equipment
  • 
Relocating the equipment with respect to the source of interference
  • 
Moving the equipment away from the device with which it is  
 
  
interfering
Adding accessories or components to a system, or modifying 
a medical device or system, may degrade the immunity 
performance. Consult qualifi ed personnel before making 
changes to the system confi guration. 
Modifi cations not expressly approved by Huntleigh Healthcare 
could void the user's authority to operate the equipment.
This device complies with part 15 of the FCC/IC Rules. 
Operation is subject to the following two conditions: 
(1) This device may not cause harmful interference, 
and (2) this device must accept any interference
received, including interference that may cause
undesired operation.
Operation of this equipment requires the prior coordination 
with a frequency coordinator designated by the FCC for
the Wireless Medical Telemetry Service.