Huntleigh Healthcare Limited Huntleigh Diagnostic Products Division SF1US 사용자 설명서

다운로드
페이지 48
 7
Saf
ety
2.1 WARNINGS
The Sonicaid™ Freedom should only be used by personnel 
familiar with the operation of electro-medical equipment, 
especially for electronic monitoring of the fetal heart rate.
A possible explosion hazard exists if used in the presence of 
fl ammable anaesthetics. 
The Sonicaid™ Freedom should not be used at temperatures 
lower than 10 or higher than 40 degrees centigrade.
Do not mount the unit directly above the patient.  Locate the 
unit so that it will not cause harm should it fall.
Do not operate the unit from the mains supply if the mains 
cable is damaged.
Do not immerse any portion of the receiver unit in water or 
other liquids. The transducers are watertight and may be used 
in water (IPX8 - TRANSDUCERS ONLY).
 If there is any damage to the transducer housings, do not use 
the transducer under water. Refer the transducer to Huntleigh 
Healthcare qualifi ed service personnel for repair.
The transducers are protected from damage if dropped. Never 
use the transducer without its protective sleeve. It can be 
removed for cleaning in accordance with the Cleaning and 
Disinfection Procedure in Section 6.
If this product is connected to another item of electrical 
equipment, it is important that the system is fully compliant 
with IEC60601-1.  
The device is generating RF-radiation. It is designed for use in 
hospitals and other clinical settings, also outside of shielded 
areas. As in other medical electrical devices, fi xed and mobile 
RF-communication devices may disrupt the performance of the 
Sonicaid™ Freedom.
The telemetry equipment is classifi ed as IIb according to 
Medical Devices Directive 93/42/EEC. The receiver unit is 
connected to the AC line without a protective earth (Class 2). 
The line voltage may be between 100 and 240V with 50 to 60Hz. 
The transducer units are powered by safe current limited low 
voltage re-chargeable batteries of 3.7V. Transducers are Class 
CF.