INTERPRETATION OF DATA
DATA QUERY -  A simple <return> sent to the unit instructs the unit to send the current 
data set. Each data element is preceded by an upper case letter for the identity of the 
following number. A typical data line is as follow: data, time, D depth, other data, B 
battery voltage and the R error code. (see the error code section for codes).
Example: 
WW660 > 
>> #000 D  74.45 T 76.4 B 5.97 R 0002
WW660 > 
This indicates that the depth is 74.45 feet, the probe temp is 76.4 F, the battery voltage 
is 5.97, and the error code is 2 meaning that the battery voltage is below 6.5 volts.
SYSTEM STATE -  A "?" command sent to the unit instructs the unit to send its current 
operating state. This will include 3 lines of data. The first line reports the model number 
and features, the software version, the unit serial number, and the user set ID number. 
The second and third lines report the applicable system settings prefixed by the 
commands used to set them as listed in the table in the section on external commands. 
Example:
WW660 > ?
Mod #WW660-077303 Ver # 6601-008  Ser #080628015  ID #001
 rn   25.0 rx 500.0 w 60.5 u0 a0 al 500.00 ah  25.00
 nh  25.00 nl 500.00  cs0681n cm0681n RTU 11 cf0  6
WW660 > 
In this example, "rn" indicates that the min range is 25.0, and the well temperature is 
set to 60.5 degrees, etc.
MODBUS COMMUNICATIONS
The RS485 port on the Well Watch is configured to communicate as a slave device on 
a two wire Modbus bus. When the RS485 Modbus port is enabled, the Well Watch 
becomes a Modbus slave device listening for data requests from a Modbus master 
device using either RTU or ASCII mode. Data requests with the address set in the Well 
  Eno Scientific  Well Watch 660DL