• Select desired volume level using “H M L” toggle switch
located in battery compartment (H = high, M = medium,
L = low)
• Select desired alarm tone using “Tone 1 Tone 2” toggle
switch located in battery compartment
• Close the battery cover and fully tighten Phillips-head
screw
• Place the monitor in protective holder
2. Wireless Alarm Transmitter (WAT-1) – Install
batteries & Adjust Settings
• To access WAT-1 battery compartment, remove case
back by fully loosening the four (4) Phillips-head screws
using the included screwdriver
• Insert 2 AAA batteries (included) into the battery
compartment ensuring correct polarity
• Attach case back and fully tighten four (4) Philips-head
screws
• Using a pen or other pointed instrument, turn on the
WAT-1 by switching the “OFF ON” toggle switch to “On”
position
• Using a pen or other pointed instrument, choose the
correct position of the “Mat Pad” toggle switch (Mat =
Floor Mat Sensor, Pad= Bed/Chair/Toilet Sensor Pad)
3. Sync/Pair WAM-1 & WAT-1, Plug-in Sensor Pad or
Floor Mat
• Press & hold WAM-1 “SYNC” button located on right
side of monitor. The red LED light will turn on indicating
that the monitor is in Sync mode.
• Press WAT-1“SYNC” button. The WAT-1 red LED light
will turn on and WAM-1will sound an audible beep.
• Both red LED lights on WAM-1 & WAT-1 will turn off. The
monitor & transmitter are now synced.
• Plug bed/chair sensor pad or floor mat sensor into the
WAT-1 “PAD PLUG”
• Always test system before each use.
4. Testing Wire-Less Patient Monitoring System
Operation
To test system, press down on the sensor pad with the
palm of your hand. The monitor will beep to indicate it
is active. After removing pressure from the sensor pad,
the monitor will immediately begin to alarm. Press the
reset button once to silence the alarm sound and put the
system into standby mode.
Instructions for Use
The WAM-1 alarm, WAT-1 transmitter and sensor pad
being used should be checked prior to each use to
ensure all are functioning properly. The “Test/Reset”
button on WAM-1 also functions as a battery tester.
Bed Sensor Pad Placement
Place Wireless Bed Sensor Pad (BEDPAD-1W) widthwise
across bed on top of mattress. The sensor pad must be
positioned on the mattress to be directly under buttocks or
shoulder of user. When placed under shoulders, the alarm
will activate when patient sits up in bed. When placed under
the buttocks, the alarm will activate when the patient vacates
the bed. Place side of pad with movement resistant cushions
down against mattress top. Mattress cover sheets should be
placed over the sensor. The WAT-1 sensor pad pocket should
hang over side of bed.
Chair Sensor Pad Placement
Place Wireless Chair
Sensor Pad (CHAIRPAD-
1W) in center of chair or
wheelchair seat directly
under buttocks of user.
If a chair cushion is used
sensor pad must be placed
on top of chair cushion
directly under buttocks of
user. The WAT-1 sensor
pad pocket should hang
over side of chair.
Low Battery Indicator Function - WAM-1 & WAT-1
1. WAM-1 Monitor
• Low battery indication for WAM-1 monitor
- WAM-1 monitor will emit an audible beep sound alert
- “Monitor Low Battery” blue LED indicator will flash
periodically
• Low battery indication for WAT-1 transmitter
- WAM-1 monitor will emit an audible beep sound alert
- “Transmitter Low Battery” yellow LED indicator will
flash periodically
2. WAT-1 Transmitter
• Yellow LED will flash three (3) times every 30 seconds
indicating that the WAT-1 transmitter is sending the low
TM
PATIENT
MONITORING SYSTEM
Continued
Personal Safety Corporation • Hiawatha, Iowa 52233 • 800.373.2873