Olympus Winter & Ibe GmbH ESG200 사용자 설명서
83
Troubleshooting
Error code Message
Cause
Remedial action
E0104
E0105
Footswitch pedal pressed
1. Release the footswitch
pedal to continue.
2. If the problem persists,
contact Olympus.
The foot switch pedal is
depressed while switching
on the electrosurgical
generator.
Release the foot switch
pedal.
Malfunction of the foot
switch.
Replace the foot switch.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0108
E0109
Handswitch pressed
1. Release the handswitch
of the instrument to
continue.
2. If the problem persists,
replace the instrument.
3. If the problem persists,
contact Olympus.
The hand switch of the
HF instrument is pressed
while switching on the
electrosurgical generator.
Release the hand switch of
the HF instrument.
Malfunction of the HF
instrument.
Replace the HF instrument.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0115
Application time limit
exceeded
1. Release the footswitch or
handswitch and reactivate
to continue.
2. If the problem persists,
contact Olympus.
The maximum time limit for
the application has been
exceeded.
Release the foot switch or
hand switch and reactivate
by pressing the foot switch
or hand switch again.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0136
Double pedal footswitch
not assigned
1. Assign the double
pedal foot switch to the
designated output socket.
2. If the problem persists,
contact Olympus.
The double pedal foot
switch is not assigned to the
corresponding output socket.
Assign the double pedal foot
switch to the corresponding
output socket. Refer to
the section “Foot switch
assignment and Autostart”
on page 41.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0140
No mode selected
1. Select a mode.
2. If the problem persists,
contact Olympus.
No mode has been selected. Select an appropriate mode.
Refer to the section “Output
settings” on page 59.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0141
Power set to zero (--)
1. Set a valid power level.
2. If the problem persists,
contact Olympus.
The output power level for
the chosen mode is set to 0.
Increase the output power
level on the Set Screen.
Refer to section “Selecting
the appropriate output
settings” on page 62.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.