Installation
WARNING
Device blackout
Be aware that the electrosurgical generator immediately turns off if the mains plug
is disconnected or if the fuses blow. During a procedure, this could result in severe
complications for the patient.
• Make sure that the mains plug is securely connected to the socket-outlet.
• Do not place portable socket-outlets on the floor to avoid trip hazards which could unplug
the cord set.
• Only use power cords as specified in the section “Specifications for the cord set” on page
91.
• Make sure that the circuit breaker is sufficient for the power requirements of the
electrosurgical generator and additional equipment.
4.3.1 Before connecting
• Make sure that the fixed socket-outlet is grounded.
Only connect the electrosurgical generator to a grounded fixed socket-outlet.
• Make sure that the fixed socket-outlet meets the power requirements as specified on the
rating plate of the electrosurgical generator.
Only connect the electrosurgical generator if the power specifications of the fixed socket-
outlet match the specifications on the rating plate.
4.3.2 Connecting
To connect the electrosurgical generator to the mains electricity proceed as follows:
1. Make sure that the electrosurgical generator is turned off, i.e. the power switch must not
be depressed.
Figure 4.1 Connecting the cord set to the electrosurgical generator
2. Connect one plug of the cord set to the power socket on the rear panel of the ESG-200.
3. Connect the other plug of the cord set to the socket-outlet of the mains electricity.
Equipotential bonding point
The purpose of additional equipotential bonding is to equalise potentials between different
metal parts that can be touched simultaneously in the medical environment. Equipotential
bonding increases electrical safety.
• Connect the pin at the rear panel of the electrosurgical generator to an equipotential
bonding lead, which should be provided by the hospital's technical department.