Clarity INT9 사용자 설명서

다운로드
페이지 35
Clarity - INT9 A/D converter 
 Installation 
12 
2.4.2 
Using multiple INT9 cards 
When using multiple A/D cards, it will be 
necessary to add the INT9 card again, 
repeating the entire procedure described in 
chapter 2.4 from Step 3
 
Fig. 6. 
System Configuration – two INT9 cards 
Individual detectors from all cards can be 
assigned to multiple instruments and 
combined together arbitrarily. In this example, 
all signals from INT7 A/D Card, as well as all 
signals from INT9 A/D Card 2 and one signal 
(Pressure) from INT9 A/D Card 1 are assigned 
to the instrument 1, while the rest of the 
signals from INT9 A/D Card 1 is assigned to 
the Instrument 2. 
While there are several cards assigned to the 
same instrument, it is necessary to select 
which card will be used for 
Start
 and 
Ready
 
signals  . 
All added A/D converters (e.g. INT9) must have 
their drivers properly installed. There must be 
correct number of items in the Windows OS