Clarity INT9 사용자 설명서
Clarity - INT9 A/D converter
Troubleshooting
29
• In the Models list select the
DataApex
Acquisition device, Model 9.00
and press the
Next
button.
• In the remaining dialogs press the
Next
button.
• Restart the PC.
• Re-check the driver in the
• Re-check the driver in the
Device Manager
as
desribed in chapter 4.3.3.
4.4 Data Acquisition – non-functional
A gray
icon, label “Disabled”, and Data Acquisition command not functional
Other manifestations of this error are also:
Method Setup - Acquisition
tab missing,
Method –
Acquisition
command non-functional,
Run
,
Stop
,
Abort
commands non-functional in the
windows
Single Analysis
and
Sequence
).
Possible causes are:
a) Detector not allocated to instrument:
Open the
System Configuration
dialog from the
main
Clarity
window using the
System -
Configuration
command and check the tab of
the corresponding instrument -
Instrument x
. If it
has no allocated detectors, add them.
Solution:
In the left-hand list
Setup Control Modules
select the correct detector that is connected to
the A/D card you are using and link it to the
corresponding instrument on the right.