DSO-1062D Digital Oscilloscope User Manual                                             7 
 
3.3.2 Use of Probe Check Wizard 
Every time you connect a probe to an input channel, you should use the probe check wizard to 
verify that this probe is operating correctly. There are two ways to do this: 
1) Use  the vertical menu (for example, push the CH1 MENU button) to set the Probe option 
attenuation factor. 
2) Press the PROBE CHECK button to use the Probe Check Wizard and  configure the probe 
option attenuation factor properly following menu prompts. 
3.4  Manual Probe Compensation 
Upon the first connection of a probe and an input channel, you  should  manually perform  this 
adjustment to match the probe to the input channel. Uncompensated or miscompensated probes 
may lead to errors or faults in measurement. To adjust the probe compensation, follow the steps 
below. 
1.  Set the Probe option attenuation in the channel menu to 10X. Set the switch on the probe to 
10X and connect the probe to Channel 1 on the oscilloscope. If you use the probe hook-tip, 
ensure  it is firmly inserted  onto the probe. Attach the probe tip to the PROBE COMP 
~5V@1KHz connector and the reference lead to the PROBE COMP Ground connector. 
Display the channel and then press the AUTOSET button. 
2.  Check the shape of the displayed waveform. 
 
3.  If necessary, use a nonmetallic screwdriver to adjust the variable capacity of your probe until 
the shape of the waveform turns to be the same as the above figure. Repeat this step  as 
necessary. See the figure below for the way of adjustment. 
Compensated correctly
 
 
 
Overcompensated 
 
 
Undercompensated