Basic Operation 
DSO-1062D Digital Oscilloscope User Manual                                             37 
5.5.4 UTILITY 
Push the UTILITY button to display the Utility Menu as follows. 
Options 
Comments 
System Info 
Display the software and hardware versions, serial number and some other 
information about the oscilloscope.     
Update 
Program   
Insert a USB disk with upgrade program and the disk icon at the top left corner is 
highlighted. Press the Update Program button and the Software Upgrade dialog 
pops up. Push F6 to upgrade or push F2 to cancel.   
Save 
Waveform 
Insert a USB disk and the disk icon at the top left corner is highlighted. Click on 
this icon and you can see the waveform pause for a while, being saved. You can 
find the saved waveform data in the USB disk. Here X represents how many 
times you press the button. Each press generates a corresponding folder. For 
example, press once and a folder 1 is generated; press twice and two folders 1, 2 
are generated.
 
Self 
Calibration 
Press this option and the Self Calibration dialog pops up. Press F6 to perform the 
self calibration. Press F4 to cancel. 
Advance 
Buzzer and time setups 
Press this button to see the dialog for buzzer and time setups. Turn the knob V0 
to select buzzer or time (with a black frame). Press V0 and the black frame turn to 
be red. Turn V0 again to set ON/OFF or time. Press V0 again to exit or save the 
setups.     
Self Calibration: The self calibration routine can optimize the precision of the oscilloscope to fit 
the ambient temperature. To maximize the precision, you should perform the self calibration once 
the ambient temperature changes by 5℃  or more. Follow the instructions on the screen.       
Tip: Press any menu button on the front panel to remove the status display and enter a 
corresponding menu. 
5.5.5 CURSOR 
Push the CURSOR button to display the Cursor Menu. 
Options 
Settings 
Comments 
Type 
Off 
Voltage 
Time 
Select a measurement cursor and display it. 
Voltage measures amplitude while Time measures time and 
frequency. 
Source 
CH1 
CH2 
MATH 
REFA 
REFB 
Select a waveform to take the cursor measurement. 
Use the readouts to show the measurement. 
Select Cursor 

S indicates Cursor 1. E indicates Cursor 2. 
A selected cursor is highlighted, which can be moved freely. Both 
cursors can be selected and moved at the same time. The box