Rigol DS2072A 2-channel oscilloscope, Digital Storage oscilloscope, DS2072A 사용자 설명서
제품 코드
DS2072A
RIGOL
DS2000A User’s Guide
5-18
Slope Trigger
Trigger on the positive or negative slope of specified time.
Trigger Type:
Press Type to select “Slope”. At this point, the trigger setting information as shown
in the figure below is displayed at the upper right corner of the screen.
Source Selection:
. The current trigger
source is displayed at the upper right corner of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Slope Condition:
Press When to select the desired slope condition.
Slope Condition:
Press When to select the desired slope condition.
: trigger when the positive slope time of the input signal is greater than
the specified time.
: trigger when the positive slope time of the input signal is lower than
the specified time.
: trigger when the positive slope time of the input signal is greater than
the specified lower limit of time and lower than the specified upper limit of time.
: trigger when the negative slope time of the input signal is greater than
the specified time.
: trigger when the negative slope time of the input signal is lower than
the specified time.
: trigger when the negative slope time of the input signal is greater than
the specified lower limit of time and lower than the specified upper limit of time.
Time Setting:
In this oscilloscope, positive slope time is defined as the time difference between the
two crossing points of trigger level line A and B with the positive edge as shown in
the figure on the next page.