Lecroy WAVEACE 2024 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2024 사용자 설명서
제품 코드
WaveAce 2024
Operator's Manual
920836 Rev A
25
ACQUISITION - TRIG_DELAY, TRDL
Description
The TRIG_DELAY command sets the time at which the trigger is to occur with respect to the
nominal zero delay position, which defaults to the center of the grid.
nominal zero delay position, which defaults to the center of the grid.
RANGE
Negative delay
: 0 to - divisions x Time/div supported by scope
Postive delay
: 0 to + division x Time/div supported by scope
If a value outside these ranges is specified, the trigger time is set to the nearest limit.
Command Syntax
TRig_DeLay <value>
<value>:=
delay time as per RANGE previously described in this topic.
NOTE: The value parameter is optional.
Query Syntax
TRig_DeLay?
Response Format
The response to the TRIG_DELAY? query indicates the trigger time with respect to the first
acquired data point.
acquired data point.
TRig_DeLay <value>
Example
The following instruction sets the trigger delay to -20 S (post-trigger):
TRDL -20S
Related Commands
TIME_DIV, TRIG_COUPLING, TRIG_LEVEL, TRIG_MODE, TRIG_SELECT,
TRIG_SLOPE
TRIG_SLOPE