Lecroy WAVEACE 2024 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2024 사용자 설명서
제품 코드
WaveAce 2024
WaveAce Remote Control
28
920836 Rev A
ACQUISITION - TRIG_SELECT, TRSE
Description
The TRIG_SELECT command selects the condition that triggers the acquisition of waveforms.
Only the Edge trigger is supported.
Only the Edge trigger is supported.
Command Syntax
TRig_Select EDGE,SR,<source>,HT,TI,HV,<hold_value>
<source>:= {C1,C2,C3,C4,LINE,EX,EX5}
<hold_value>:=
Refer to the trigger information provided for your oscilloscope model.
NOTE: When specifying <hold_value>, the unit S (seconds) is optional.
Query Syntax
TRig_SElect?
Response Format
TRig_SElect <trig_type>,SR,<source>,HT,<hold_type>,HV,<hold_value>
HV2 only returned if <hold_type> is P2 or I2.
Example
The following instruction selects the single-source trigger with Channel 1 as trigger source.
Hold type and hold value are chosen as "pulse smaller" than 20 ns:
Hold type and hold value are chosen as "pulse smaller" than 20 ns:
TRSE SNG,SR,C1,HT,PS,HV,20 NS