Lecroy WAVEACE 2024 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2024 사용자 설명서
제품 코드
WaveAce 2024
Operator's Manual
920836 Rev A
33
CURSOR - CURSOR_SET, CRST
Description
The CURSOR_SET command allows you to position any one of the independent cursors at a
given grid location. When you are setting a cursor position, you must specify an active trace
relative to which the cursor is positioned.
given grid location. When you are setting a cursor position, you must specify an active trace
relative to which the cursor is positioned.
The CURSOR_SET? query retrieves the current position of the cursor(s). The values returned
depend on the grid type selected:
depend on the grid type selected:
HREF Horizontal Cursor A
VREF Vertical Cursor A.
HDIF Horizontal Cursor B
VDIF Vertical Cursor B.
Command Syntax
<trace>:CuRsor_SeT <cursor>,<position>,...
<trace>:= {TA,TB,TC,TD,C1,C2,C3,C4,M1,M2,M3,M4,M5,M6,M7,M8,M9,M10}
<cursor>:= {HABS, VABS, HREF, HDIF, VREF, VDIF}
<position>:= 0 to 10 DIV (horizontal);-3.99 to 3.99 DIV (vertical)
Query Syntax
<trace>:CuRsor_SeT? <cursor>
<cursor>:= {HREF, HDIF, VREF, VDIF}
No <cursor> implies ALL, even though not all are visible.
Response Format
<trace>:CuRsor_SeT <cursor>,<position>,...<cursor>,<position>
If the position of a cursor cannot be determined, its position will be given as
UNDEF
.
Example
The following instruction positions the VREF and VDIF cursors at +3 DIV and -2 DIV respectively,
using Trace TA as a reference:
using Trace TA as a reference:
TA:CRST VREF,3DIV,VDIF,-2DIV
Related Commands
CURSOR_MEASURE, CURSOR_VALUE?, PARAMETER_VALUE?