To Perform a Functional Check
Perform this quick functional check to verify that the instrument is operating
correctly.
correctly.
1. Turn on the instrument.
Use the power cord designed for the oscilloscope only.
Use a power source that delivers 100 to 240 V, 45Hz to 440Hz.
Turn on the instruments, and wait until the display shows the waveform window.
Push the Storage
button, select Storage in the top menu box and push the
Factory menu box, then continue pressing Load.
Figure 1-4
Turn on the instrument
WARNNING:
To avoid electric shock, be sure the oscilloscope is properly grounded.
To avoid electric shock, be sure the oscilloscope is properly grounded.
2.
Input a signal to a channel of the oscilloscope
①
Set the attenuation switch on the probe to 10X and connect the probe to Channel
1 on the oscilloscope.
To do this:
To do this:
Align the slot in the probe connector with the socket on the Bayonet Nut
Connector (BNC) connector of CH1 or CH2.
Push to connect, and twist to the right to lock the probe in place.
Attach the probe tip and ground lead to the PROBE COMP connector.
!
Storage button
Power