Application example: measure the phase deviation of the input signals of two 
channels. 
 
Method 1: Use Lissajous method 
 
1.  Connect a sine signal to CH1 and then connect a sine signal with the same 
frequency and amplitude but a 90° phase deviation to CH2. 
2.  Press AUTO and adjust the vertical position of CH1 and CH2 to 0. 
3.  Press X-Y to select “CH1-CH2”, rotate Horizontal    SCALE to ajdut the 
sample rate properly to get better Lissajous figure for better observation and 
measurement. 
4.  Roate VERTICAL    POSITION to adjust CH1 and CH2 to get better 
observation of signals. At this point, the circle as shown in the figure below 
should be displayed. 
 
 
 
5.  As shown in the figure above, the distances from the crossing points of axis and 
the circle to the origin of the coordinates are approximately equal. Thus, the 
phase deviation angle θ=±arcsin1=90°
 
Note:
 
  In YT mode, the oscilloscope could use any sample rate (within the guaranteed 
range) to capture waveform. The maximum sample rate of XY mode is 500 
MSa/s. Generally, reducing the sample rate properly could improve the display 
effect of Lissajous figure.   
  When XY mode is enabled, 
 will be disabled automatically. 
  Press X-Y to select “CH1-CH2, CH1-CH3, CH1-CH4, CH2-CH3, CH2-CH4,