Rigol DS1074B 4-channel oscilloscope, Digital Storage oscilloscope, DS1074B 사용자 설명서

제품 코드
DS1074B
다운로드
페이지 164
RIGOL 
©  2008 RIGOL Technologies, Inc. 
User‟s Guide for DS1000B Series 
 
2-17 
 
Selecting an FFT Window 
DS1000B series oscilloscope provides four FFT windows. Each window is a trade-off 
between  frequency  resolution  and  amplitude  accuracy.  It  depends  on  the  desired 
measurement and the source signals characteristics to determine the window to use. 
Use the following guidelines to select the best window.   
Table 2-9 FFT Windows 
Window 
Features 
Best for measuring 
Rectangle 
Best frequency resolution 
and worst amplitude 
accuracy. 
 
This is essentially the same 
as no window. 
 
Transients  or  bursts,  the  signal 
levels  before  and  after  the  event 
are nearly equal. 
Equal-amplitude  sine  waves  with 
fixed frequencies. 
Broadband random noise with a 
relatively slow varying spectrum. 
Hanning 
 
 
 
Hamming 
Better frequency resolution, 
poorer amplitude accuracy 
than Rectangular. 
 
Hamming has slightly better 
frequency  resolution  than 
Hanning. 
Sine, periodic, and narrow-band 
random noise. 
 
Transients  or  bursts  where  the 
signal  levels  before  and  after  the 
events are significantly different. 
Blackman 
Best amplitude accuracy, 
worst frequency resolution. 
Single  frequency  waveforms,  to 
find higher order harmonics. 
 
Key points: 
FFT Resolution: 
The FFT resolution is the quotient between sampling rate and 
the number of FFT points. With a fixed FFT points, the lower sampling rate, the 
better the resolution. 
Nyquist Frequency 
The  highest  frequency  that  any  real-time  digitizing  oscilloscope  can  acquire 
without aliasing. It‟s normally half of the sample rate. This frequency is called 
the  Nyquist frequency.  Frequency  above  the  Nyquist  frequency  will  be  under 
sampled, causing a situation known as aliasing.