G
ETTING 
S
TARTED 
M
ANUAL
 
43 
 WSMXs-B-GSM-E Rev A
 
T
RIGGERING
 
Overview 
The oscilloscope uses many waveform capture techniques that trigger on features 
and conditions that you define. These triggers fall into two major categories: 
Edge activated by basic waveform features or conditions such as a positive or 
negative slope, and hold-off  
SMART Trigger® sophisticated triggers that enable you to use basic or complex 
conditions for triggering. 
Use Edge Triggers for simple signals, and the SMART Triggers for signals with rare 
features, like glitches. 
This Getting Started Manual contains information on setting up the Edge Trigger. 
Reference the online Help for information on setting up and using SMART 
Triggers. 
Trigger Terms 
 
 
Trigger Delay - the amount of time that the oscilloscope trigger is adjusted 
from a zero value. You can adjust the oscilloscope’s trigger position from 0-
100% pre-trigger (left to right on the grid), or from 0-10,000 divisions (in time 
units) of post-trigger.  
 
Coupling – the type of signal coupling at the input of the trigger source: AC, 
DC, LFREJ, HFREJ 
 
Level – the threshold at which the trigger will occur, specified in volts.  
 
Slope – the direction of the trigger voltage transition used for generating a 
particular trigger.