OPERATION 
43 
 
4.  For ACW and DCW tests, turn the 
scroll wheel to set the desired test 
voltage.  
 
 
 
Range 
0.1kV ~ 5.0kV(ACV)/6.0kV(DCV) 
 
5.  For IR tests, turn the scroll wheel 
to set the voltage to one of five 
settings.  
 
 
 
Range 
50V, 100V, 250V, 500V, 1000V 
 
 
     
 
 
 
H = 2 2 . 0 m A  
 
 
 
 
m A
L
0
=
0 0
.
   
7 4 S  
 
 
 
 
 
= 1
T
       
   
CW
D
A CW
R  
I
D
O E
M
T M E
I
Voltage setting
 
1 00
 
k V
A
E
R
 
D
 
Y
A
m
 
 
6.  Press the START button. The test 
starts automatically and the tester 
goes into the TEST status. 
START
 
 
7.  The TEST indicator will be lit 
orange when in the TEST status. 
TEST
 
 
8.  The test will start by showing the remaining 
test time. The test will continue until the test is 
finished or the test is stopped.  
 Note 
ACW and DCW mode can also edit the test voltage 
when a test is running. 
Conversely, IR tests cannot edit the test voltage 
when a test is running.