Voltcraft RGB-2000Tester 122953 사용자 설명서

제품 코드
122953
다운로드
페이지 50
22
RS232 interface
The measuring device is fitted with a serial interface (11) for exchanging data with a computer. This can be
found on the right hand side under a cover. Open the cover with a pointed object. The interface is in the form
of a 3.5mm jack plug socket and requires a compatible data cable, which is an optional extra.
You can order the matching interface cable using the following order numbers:
Order no.: 122156 RS232 interface cable with jack plug and 9-pole D-USB plug
Order no.: 122703 USB interface cable with jack plug and USB plug
The data cable has the following components:
Jack plug 3.5 mm mono
9-pole D-SUB socket for PC
Center contact 
Pin 4
External contact 
Pin 2
A 2.2 KOhm is required between pin 2 and pin 5.
The serial data signal is made up of 16 Bits in the following order:
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
Each data bit has the following function:
D0
End signal
D1 to D8 Measuring value
D1 = 
smallest digit (LSD), D8 = largest digit (MSD).
If 1234 is displayed, the bit rate is: „00001234“
D9
Decimal point (DP) at the corresponding spot from right to left; 
0= no DP; 1=1DP; 2=2DP; 3=3DP
D10
Polarity; 0 = positive, 1 = negative
D11+D12 Measuring value on the display; R = 70 / G = 71 / B = 72 / H = 42 /  S = 73 / L = 99 
D13
Data bit when sending the colour value: Red = 1 / Green = 2 / Blue = 3 / Hue = 4 / Sat = 5 / Lum = 6
D14
4
D15
Start bit = 02
The RS232 data format is: 9600, N, 8, 1
Maintenance and cleaning
The colour analysis device RGB-2000 is maintenance-free apart from replacing the batteries and cleaning it
once in a while. Use a clean, lint-free, antistatic and dry cloth to clean the device. Do not use any abrasive or
chemical agents or detergents containing solvents.
Battery replacement
If the battery symbol 
appears on the display, the battery has to be replaced as soon as possible to pre-
vent wrong measuring results.