Texas Instruments Advanced bqMTester Multi-Station Test and Program Board Texas Instruments ADVANCED-BQMTESTER ADVANCED-BQMTESTER 데이터 시트

제품 코드
ADVANCED-BQMTESTER
다운로드
페이지 62
Appendix G
(i) bqTester.RES
3. Tool Box Objects
In addition to the standard Visual Basic GUI objects, the tool box also contains the
following objects:
(a) Bq80xRW1
(b) clsPassword
(c) clsReminder
(d) EV2x00ID
(e) TesterLog
(f) GlobalConfig
4. frmMainS
Important functions
(a) Sub SetState(nState As StatusType, Optional sStatusText As String = "")
Sets the status of the
program. The status types are:
(i) vb_cbReady = 1
(ii) vb_cbbusy = 2
(iii) vb_cbPASS = 4
(iv) vb_cbfail = 8
(v) The status type is defined in the Class Module SingleSite.
(b) cmdStart_Click()
This performs Tester operations in sequence.
(i) Indirectly causes the MainProcess() in SingleSite to be executed.
5. SingleSite Class
Important functions
(a) MainProcess()
This controls program flow.
(b) Note: Implement your changes in CustomFunctionX() that is defined in modCustom.bas.
(c) Program Flow:
(i) Power up bqMTester board.
(ii) Get the Device and version from the Target device.
(iii) Call CustomFunction1.
(iv) Write Data Flash Image in ROM mode.
(v) Wait for Gas gauge to execute startup code.
(vi) Call CustomFunction2.
(vii) Do Calibration.
(viii) Post Calibration checks.
(ix) bq8060 voltage calibration
Has to be done in GG mode for this part.
(x) Write Manufacturer Data.
(xi) Write Date.
(xii) Call CustomFunction3.
(xiii) Write Serial Number
Everything passed!!!
(xiv) Put device into ROM mode before powering off.
(xv) Wait for any pending packets to go out before turning off power.
(xvi) Cal board power off.
(d) Error Processing for Program Flow
(i) Flag event current status.
(ii) Clear all errors.
(iii) Flag event serial number.
(iv) Send GG to ROM.
(v) Wait for any pending packets to go out before turning off power.
(vi) Cal board power off
6. InitializeModules()
(a) Initialize communication functions
modCommFunc.
(b) Initialize calibration board
modCalBoard
(c) Initialize configuration data
modDelays
(d) Initialize custom code
modCustom
7. TerminateModules()
(a) Terminate custom code
modCustom
(b) Terminate configuration data
modDelays
(c) Terminate communication functions
modCommFunc
(d) Terminate the main process
modComm
46
Programming Guide
SLUU397A
February 2011
Revised May 2011
Copyright
©