Spectrum Analyzer Software Reference
4.2
Traces
All measurement data is stored in traces. A total of four traces are available and a function or mode can
be assigned to each of them in the Traces tab (see
).
Figure 13. Traces Tab, Trace 0
Four modes are available and define the function of each trace.
ACT (Actual) gives the current measured value
MAX (Maximum) shows the maximum value measured per frequency point since the last measurement
start button press.
AVG (Average) shows an average value per frequency point since the last measurement start.
OFF removes the trace from the graph
For the average function a sliding weighted average function is implemented. The formula looks like: AVG
= (5*AVG[last measurement] + 1*ACT)/6.
Starting a measurement by pressing the Start button sets AVG and MAX to ACT values. In Single Sweep
mode all traces will show the same result.
Pressing the Clear Button resets the average and maximum trace, so MAX or AVG will be identical with
next ACT trace measurement.
The Hold button will stop updates to this trace. It will not stop the measurement
it only stops updating
data on a selected trace. This feature is Useful to compare for example two devices or settings.
Each trace can be exported to a comma separated text file (csv) with frequency and dBm values. This is
useful to store or share measurement results. Good file names are key.