Texas Instruments LMP8480 and LMP8481 Evaluation Tool LMP8480MEVM-S/NOPB LMP8480MEVM-S/NOPB 데이터 시트

제품 코드
LMP8480MEVM-S/NOPB
다운로드
페이지 17
 
 
 
March 2013 
 
 
 
 
 
 
  
Literature Number SNOU031A 
 
 
 
 
 
 
 
  
 
 
5.  Board Test Procedure for Bi-Directional Devices (LMP8481-S) 
 
Required Equipment: 
 
•  1 ea.  10V, >1A power supply 
•  2 ea.  4.5 digit or better DMM (Agilent 34401 or equiv) 
•  1 ea.  20 ohm, >20W power resistor or programmable load sinking 0.5A. 
•  2 ea  0.1” Jumper blocks 
•  Misc test leads for connections 
 
Connect board as shown, but do not connect SENSE- to load yet. 
 
1.  Connect 10V power supply as VSOURCE (turn off power for now). 
2.  Connect DMM1 to the output and ground terminals (+ to output, - to GND) 
3.  Connect DMM2 to the TP1 terminals or sense terminals (+ to RSP, - to RSN) 
4.  Place jumpers across JP1 2&3 and JP2 2&3 to set reference to VS/2. 
5.  Confirm SENSE- (load) is not connected. 
6.  Turn on power and record DMM1 reading (should be 5V* - this is the VREF voltage). 
7.  Now connect RLOAD to SENSE- as shown. 
8.  Record reading from DMM1 (should be around 8V for the “S” device) 
9.  Record reading from DMM2 (should be around 50mV) 
10.  The difference in the two readings of DMM1 should be within 0.1% of device gain (60) times the 
DMM2 reading plus the VREF reading in step 6. 
11.  Reverse the SENSE+ and SENSE- leads. DMM2 should measure -100mV and the output should be 
at 3V for the “S” device. 
12.  Turn off Vsource or disconnect RLOAD to prevent excessive heating. 
 
-
+
RLOAD
VSOURCE
DMM 1
DMM 2
-+
+
-
 
Figure 9: Setup for Bi-Directional Devices