Figure 11 - Results for Shorted Input Test without Calibration 
 
F.  Capturing Data with Calibration 
 
1.  In the “Measurement” tab, go to “Quick Control  BGCAL_MODE” and change the 
background calibration to “001: Offset Cor / Gain Est”. 
2.  Click on the “Run” button again to view the output voltage results.  A mean output reading 
closer to 0V should be plotted similar to Figure 12.  This decrease in the mean output reading 
demonstrates the LMP90100 offset calibration feature.