B.  Installing/Opening the Software – skip this step if it’s already done.  If not, follow section 9.0 
to install and open the LMP90100 Sensor AFE software. 
 
C.  Connecting and Powering the Boards 
1.  Connect the LMP90100EB to the SPIO-4 board as seen in Figure 4.  
2.  Connect SPIO-4 board to a PC via USB. 
3.  Use a multimeter to measure LMP90100EB’s JP6 and JP7; they should all be approximately 
5.0V.  If they are not, check your power supplies and jumpers.  
 
 
D.  Connecting the Sensor to the LMP90100EB 
1.  Connect a 3-wire RTD to J10 as seen in the image below.  The white wire should be at 
J10.P1, and the red wires should be at J10.P3 and J10.P4. 
 
 
Figure 15 – Jumper Settings (Default) for the 3-wire RTD Example 
 
 
E.  Configuring the LMP90100 Using the Sensor AFE Software 
 
Follow the step-by-step instructions under the “HelpBar” mini-tab (left hand side of the GUI) to 
configure the LMP90100 for this example.  These step-by-step instructions are discussed in 
details below, and the recommended configuration should look similar to Figure 16.