Texas Instruments PCM3070RHBEVM-K Evaluation Module PCM3070RHBEVM-K PCM3070RHBEVM-K 데이터 시트

제품 코드
PCM3070RHBEVM-K
다운로드
페이지 37
PCM3070 Control Software
Figure 9. Audio Inputs Panel
Before changing a control, see the data sheet to ensure that a particular control is compatible with the
current state of the codec. As an example, some controls in the Analog Setup panel must be modified in
a particular order as described in the data sheet. Other controls must only be modified with a specific
hardware setup, such as powering up the AVDD LDO.
All controls update their status with respect to the register contents in the following conditions:
A panel is opened.
The Execute Command Buffer button in the Command-Line Interface is pressed.
The Refresh button at the bottom right of a panel is pressed.
4.2
Status Flags Panel
The PCM3070 status flags can monitored in the Status Flags panel (
which is located below
the Panel Selection Tree . Pressing the POLL button continuously reads all the registers relevant to each
flag and updates those flags accordingly. The rate at which the registers are read can be modified by
changing the value in the Polling Interval numeric control. Note that a smaller interval reduces
responsiveness of other controls, especially volume sliders, due to bandwidth limitations. By default, the
polling interval is 200 ms and can be set to a minimum of 20 ms.
The Sticky Flags tab contains indicators whose corresponding register contents clear every time a read is
performed to that register. To read all the sticky flags, click the Read Sticky Flags button.
16
PCM3070EVM-K Evaluation Module
SLAU331A – January 2011 – Revised October 2012
Copyright © 2011–2012, Texas Instruments Incorporated