Rigol DM3058E Digital-Multimeter, DMM, DM3058E 데이터 시트
제품 코드
DM3058E
RIGOL
Programming Guide for DM3058/DM3058E
6-17
Example 8: Querying the Error Information
The status register provided enables user to diagnose the current error operation on
the basis of its query function, searching the error reason. For the detailed commands,
refer to Chapter 2. As the commands system of status register is in common use in all
RIGOL, Fluke and Agilent, take RIGOL commands for example.
The programming implementations——RIGOL Commands:
0 *RST
/* Reset the instrument */
1 cmdset rigol
/* Select Agilent commands */
2 *cls
/* Clear the instrument state */
3 :measure?
4 :measure?
4 :measure?
/* Beeping */
5 SYST:ERR?
6 Return “query interrupted”
7 *esr?
8 Return 4
9 *cls
10 **cls
6 Return “query interrupted”
7 *esr?
8 Return 4
9 *cls
10 **cls
/*Beeping*/
11 SYST:ERR?
12 Return “syntax error”
13 *esr?
14 Return 32
15 *cls
16 cmdset
12 Return “syntax error”
13 *esr?
14 Return 32
15 *cls
16 cmdset
/*Beeping*/
17 SYST:ERR?
18 Return “Parameter error”
19 *esr?
20 Return 16
21 *cls
22 SYST:ERR?
18 Return “Parameter error”
19 *esr?
20 Return 16
21 *cls
22 SYST:ERR?
/* Read data forcibly and emit beep once */
23 Return “Query Unterminated”
24 *esr?
24 *esr?