Measuring Frequency 
XWWarning 
To avoid electrical shock, disregard 
the bar graph for frequencies >1 kHz. 
If the frequency of the measured 
signal is >1 kHz, the bar graph and 
Z 
are unspecified. 
AC Voltage Frequency
Hz
V
Hz
 
edy09f.eps 
The Meter measures the frequency of a signal by counting 
the number of times the signal crosses a trigger level each 
second. The trigger level is 0 V for all ranges. 
1.
  Press g to turn the frequency measurement 
function on and off. Frequency works with ac volts only. 
2.
  In frequency, the bar graph and range annunciator 
indicate the AC voltage present. 
3.
  Select progressively lower ranges using manual ranging 
for a stable reading. 
Making Low Impedance Capacitance 
Measurements  
For making capacitance measurements on cables with 
ghost voltage, hold 
qwhile turning on the Meter to 
switch the Meter into LoZ, (low input impedance) 
Capacitance mode. In this mode, capacitance 
measurements will have a lower accuracy and lower 
dynamic range. This setting is not saved when the Meter is 
turned off or goes into sleep mode.