Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 데이터 시트

제품 코드
LCR-816
다운로드
페이지 91
 
8
Set Nomval 
PC send command   
SORT
:
NOMV
 
+32.0000
<
 ^END^M
> or 
<^J^M>
 
ASCII CODE  = 
53 4F 52 54
 3A 
4E 4F 4D 56
 20 
2B 33 32 2E 30 30 30 30
 0A 0D  (Hex format)   
LCR Meter Response 
SORT
:
NOMV  
32.0000
<
^END
ASCII CODE  = 
53 4F 52 54
 3A 
4E 4F 4D 56
 20 
20 33 32 2E 30 30 30 30
 0A  (Hex format)   
PC send command 
SORT
:
NOMV
 -
32.0000
<
 ^END^M
> or 
<^J^M>
 
ASCII CODE  = 
53 4F 52 54
 3A 
4E 4F 4D 56
 20 
2D 33 32 2E 30 30 30 30
 0A 0D  (Hex format)   
LCR Meter Response   
SORT
:
NOMV -
32.0000
<
^END
ASCII CODE  = 
53 4F 52 54
 3A 
4E 4F 4D 56
 20 
2D 33 32 2E 30 30 30 30
 0A  (Hex format)  
 
Set Voltage 
PC send command   
MAIN
:
VOLT
 
1.000
<
 ^END^M
> or 
<^J^M>
 
ASCII CODE  = 
4D 41 49 4E
 3A 
56 4F 4C 54
 20 
31 2E 30 30 30 
0A 0D  (Hex format)   
LCR Meter Response   
MAIN
:
VOLT
 
1.000
<
^END
ASCII CODE  = 
4D 41 49 4E
 3A 
56 4F 4C 54
 20 
31 2E 30 30 30 
0A  (Hex format)   
Set Average 
PC send command   
STEP
:
AVER
 
1.00
<
 ^END^M
> or 
<^J^M>
 
ASCII CODE  = 
53 54 45 50
 3A 
41 56 45 52
 20 
31 2E 30 30 
0A 0D  (Hex format)   
LCR Meter Response   
STEP
:
AVER
 
1.00
<
^END
ASCII CODE  = 
53 54 45 50
 3A 
41 56 45 52
 20 
31 2E 30 30 
0A  (Hex format)   
Store Memory 
PC send command   
MEMO
:
STOR
 
1.00
<
 ^END^M
> or 