Benning DUTEST pro Multi-Tester 050156 데이터 시트

제품 코드
050156
다운로드
페이지 6
09/ 2014 
DUTEST
® 
pro 
 
screwdriver and push off the battery compartment 
cover  in downward direction at the same time. 
-  Replace the exhausted batteries by three new 
ones of type AA (LR06). Make sure that the new 
batteries are inserted with correct polarity!
-  Push the battery compartment cover  back onto 
the housing until the catch  locks into place with 
an audible click.
Note:
The battery compartment cover  is provided with an 
integrated magnet and a belt clip for attachment of the 
device.
10.  Technical data:
-  Built and tested according to: DIN EN 61010-1 
and -031, IEC 61010-1 and -031, DIN EN 62471
-  Protection against external voltages: max. 
400 V
, 50 Hz / 60 Hz
-  Measuring category: CAT III 300 V to earth
-  Continuity test:
acoustically by means of buzzer for measuring 
resistances R ≤ 100 Ω to 200 Ω
visually by means of three LED steps:
LED 100 Ω for measuring resistances R ≤ 100 Ω
to 200 Ω
LED 1 kΩ for measuring resistances R ≤ 1 kΩ
LED 10 kΩ for measuring resistances R ≤ 10 kΩ
-   Range of tolerance: ± 20 % of the maximum value 
of the LED step
-  Buzzer volume: step 1 (low), step 2 (medium), 
step 3 (high) and step 4 (buzzer: OFF, LED torch: 
ON).
-  Open-circuit voltage: ≤ 4.5 V
-  Testing current: ≤ 30 µA
-  Diode test: ~ 1.5 V, max. 30 µA
-  External-voltage detection: 6 V - 400 V AC / DC, 
50 Hz / 60 Hz
-  Internal resistance: 166 kΩ
-  Current consumption: I
s
 < 3.5 mA (400 V)
-  Polarity indication: from ± 6 V on 
-  External conductor test (phase): U
n
≥ 230 V, 50 Hz
/ 60 Hz 
-  Cable break detector: U
n
≥ 230 V, 50 Hz / 60 Hz
-  Contamination level: 2
-  Protection category: IP 40 (DIN VDE 0470-1 IEC / 
EN 60529) 
4 – first index: protection against access to dan-
gerous parts and protection against solid impuri-
ties of a diameter > 1.0 mm
0 – second index: no protection against water 
-  Battery: 3 x 1.5 V batteries of type AA (LR06) 
-  Weight: approx. 130 g
-  Measuring line with test handles: approx. 
1,000 mm 
-  Operating temperature range: - 15 °C to + 55 °C 
-  Storage temperature range: - 15 °C to + 55 °C 
-  Relative air humidity: 20 % to 80 % 
-  Luminosity of the LED torch:
step 1 (25 %), step 2 (50 %), step 3 (75 %), step 4 
(100 %)
11.  General maintenance
Clean the exterior of the device with a clean dry cloth. 
If there is contamination or deposits in the area of the 
battery or the battery housing, clean these areas as 
well by means of a dry cloth.
If the device is stored for a longer period of time, 
remove the batteries from the device!
12.  Environmental protection 
At the end of product life, dispose of the 
unserviceable device as well as used batte-
ries via appropriate collecting facilities provi-
ded in your community.