Emissivity Adjustment 
The correct emissivity values are important for you to make the most accurate 
temperature measurements. Emissivity of a surface can have a large effect on 
the apparent temperatures that the Imager observes. Understanding the 
emissivity of the surface being inspected can, but may not always, allow you to 
obtain more accurate temperature measurements. 
If you set a value that is <0.60,  shows on the Imager display with this 
caution: 
Caution: 
Emissivity <0.6 
(see Manual) 
Note 
Surfaces with an emissivity of 
<0.60 make reliable and consistent 
determination of actual temperatures problematic. The lower the 
emissivity, the more potential error is associated with the Imager's 
temperature measurement calculations. This is also true even when 
adjustments to the emissivity and reflected background adjustments 
are performed properly. 
Emissivity is set directly as a value or from a list of emissivity values for some 
common materials. 
Note 
If the Display is set to Display All, you see the information about 
current emissivity as 
ε = x.xx
Adjust by Number 
To set the emissivity value: 
1.  Push . 
2.  Push / to highlight Measurement
3.  Push  or  to view the menu. 
4.  Push / to highlight Emissivity