performance verification test procedures. Alternative equipment may be used as long as the accuracy is as good 
as or better than the specifications listed.
Standard Source
Operating Range
Accuracy 
Required
Recommended 
Equipment
K-Type Thermocouple 
Simulator or Calibrator
–50 ~ 1000 °C
–58 ~ 1832 °F
≤  ±0.1% + 1 °C
CL-300-1000C from OMEGA or Fluke-5520A
Multimeter
DC 1000.0 mV
≤ ±0.1%
Agilent-U1251A, U1252A or Agilent-34405A
Ice Point Reference 
Chamber
–5 ~ 38 °C
≤ ±0.1% + 0.5 °C
TRCIII from OMEGA or Fluke -5520A
Adjustment Procedures
The following procedures are to be performed in sequence after another to ensure a correct calibration sequence.
 
Celsius Temperature Offset Adjustment
 
1. 
Set the operating range of the digital multimeter to DC 1000.0 mV.
 
2. 
Set the reference standard to 0 ºC.
 
3. 
Switch the adapter’s slider to the ºC position.
 
4. 
Connect the K-type miniature connector from the reference standard into the input of U1586B adapter.
 
5. 
Plug in the banana ends of the adapter’s output to the V and COM input terminals of the digital 
                     multimeter.
 
6. 
Wait for at least 30 seconds for the signal conversion from ºC to Volts to stabilize at the multimeter 
                    display.
 
7. 
Observe that the multimeter reading should display a voltage of 0.0 mV ± 0.2 mV.
 
8. 
If the multimeter reading display does not meet the required voltage, unscrew 2 Phillips head screws 
                     at the back panel of the adapter using a Phillips tip screwdriver and remove the rear cover.
 
9. 
Look for VR3 (refer to Figure 1 for the location of VR3 in the adjustment diagram) and carefully adjust 
                     the variable resistor using a small Phillips tip screwdriver until the display of the multimeter meets 
                     the required reading of 0.0 mV ± 0.2 mV.
 
10.  After performing the adjustment of VR3, remember to replace the rear cover of the adapter or proceed 
                    to the next calibration process.
 
 
Celsius Temperature Gain Adjustment
 
1. 
Set the operating range of the digital multimeter to DC 1000.0 mV.
 
2. 
Set the reference standard to 300 ºC.
 
3. 
Switch the adapter’s slider to the ºC position.
 
4. 
Connect the K-type miniature connector from the reference standard into the input of U1586B adapter.
 
5. 
Plug in the banana ends of the adapter’s output to the V and COM input terminals of the digital 
                     multimeter.
 
6. 
Wait for at least 30 seconds for the signal conversion from ºC to Volts to stabilize at the multimeter 
                     display.
 
7. 
Observe that the multimeter reading should display a voltage of 300.0 mV ± 0.2 mV.
 
8. 
If the multimeter reading display does not meet the required voltage, unscrew 2 Phillips head screws 
                     at the back panel of the adapter using a Phillips tip screwdriver and remove the rear cover.
 
9. 
Look for VR1 (refer to Figure 1 for the location of VR1 in the adjustment diagram) and carefully adjust 
                     the  variable resistor using a small Phillips tip screwdriver until the display of the multimeter meets 
                     the required reading of 300.0 mV ± 0.2 mV.
 
10.  After performing the adjustment of VR1, remember to replace the rear cover of the adapter or proceed 
                    to the next calibration process.
 
Fahrenheit Temperature Offset Adjustment
 
1. 
Set the operating range of the digital multimeter to DC 1000.0 mV.
 
2. 
Set the reference standard to 32 ºF.
 
3. 
Switch the adapter’s slider to the ºF position.
 
4. 
Connect the K-type miniature connector from the reference standard into the input of U1586B adapter.
 
5. 
Plug in the banana ends of the adapter’s output to the V and COM input terminals of the digital 
 
 
multimeter.
 
6. 
Wait for at least 30 seconds for the signal conversion from ºC to Volts to stabilize at the multimeter 
 
 
display.
 
7. 
Observe that the multimeter reading should display a voltage of 32.0 mV ± 0.2 mV.
 
8. 
If the multimeter reading display does not meet the required voltage, unscrew 2 Phillips head screws 
                    at the back panel of the adapter using a Phillips tip screwdriver and remove the rear cover.
 
9. 
Look for VR4 (refer to Figure 1 for the location of VR4 in the adjustment diagram) and carefully adjust 
                     the variable resistor using a small Phillips tip screwdriver until the display of the multimeter meets 
                     the required reading of 32.0 mV ± 0.2 mV.
 
10.  After performing the adjustment of VR4, remember to replace the rear cover of the adapter or proceed 
                     to the next calibration process.