Instruction Manual
EL 9000 A Series
Date: 11-16-2011
The adjusted rise/fall time is still effective here, but the pul-
se widths are now determined by the trigger signal, which 
is fed into the trigger input. The trigger signal must be 
square wave, for levels see „8. The analogue interface“. 
As long as external trigger is active the pulse times of A 
and B are not indicated in the display and are also not 
adjustable anymore. The display indicates the condition 
as „
Ext. trigger
“.
The rotary knob 
Selection(5) is used to select the set 
values  for A  and  B,  as  well  as  both  pulse  times,  and 
Setting(6) is used to change them. The display shows 
the letters 
A
 and 
B
to indicate the assignment. When 
navigating between the set values, an arrow (->) is placed 
in front of the currently selected one. A rise/fall time for 
the pulsed operation is also adjustable. Both, rise and fall 
time, are identical and can‘t be set seperately for A and 
B. It is displayed as slope in the format ∆value/time, like 
for example in figure 10 with 100A/20ms.
The 
Level A/B operation is started as soon as the load 
input is set to on.
Note: during Level A/B operation following applies: the 
set value of A must always be greater than or equal to 
B. Thus you can adjust A downwards only to the value 
of B and B upwards only to the level of A. If it seems that 
Level A is not adjustable somehow this might be caused 
by Level B being the same value.
Figure 11
6.7.2  Level B
When  switching  to  „Level  B“  with  the  selector 
Level
Control(3), the set values of Level B become active. This 
mode works the same way as 
Level A.
In 
Level B mode, the load can be switched to remote 
control from a PC and via the interface, in order to control 
and monitor it like during manual operation.
When  changing  to  remote  control  the  currently  selec-
ted 
Level Control setting is retained and can then be 
changed  by  a  command,  except  with  GPIB  interface 
IF-G1.
6.7.3  Level A/B (pulsed operation)
This mode combines two set values A and B with two 
seperately  adjustable  pulse  widths  for A  and  B.  The 
electronic load generates automatic set value leaps bet-
ween A and B with those values. The rise/fall time is also 
adjustable. This pulsed operation is only applied to the 
set value which belongs to the selected MODE, means 
in CV the voltage is affected etc. The other set values 
remain constant. Also see figures 11 and 12.
The pulse width of A is dedicated to the set value of A etc. 
The sum of the pulse widths results in a period t, which 
represent a certain frequency f=1/t. The pulse widths are 
adjustable  from  50µs...100s  which  results  in  a  period 
of  100µs...200s,  which  corresponds  to  a  frequency  of 
10kHz...0.005Hz.
Note: Alarms like OVP or PF (power fail) (see section 6.1, 
item „Alarm management“) which switch off the input also 
stop the pulsed operation. It can be resumed as soon as 
all alarms are gone and have been acknowledged.
External trigger
The external switchover between A and B, realised with 
the trigger input on the analogue interface, is only availab-
le in Level A/B mode. The trigger input has to be activated 
in the setup menu with the option 
Trigger mode
Device configuration“. The default setting is 
internal
. By 
setting it to 
external
 the switchover between A and B can 
only be done via the trigger input.
Handling the device 
Figure 10. Level A/B operation with pulse width adjustment
Figure 9. Normal load operation in CP regulation mode