Instruction Manual
EL 9000 A Series
Date: 11-16-2011
Remote control
8.  The analogue interface
Introduction
The analogue interface is a 15pole Sub-D socket and is 
located at the rear side. It is designed to remotely control 
the  most  important  functions  of  the  electronic  load  by 
external hardware (eg. SPS, switches, relays) with it. 
The load requires to be switched to external control
in order to use the analogue interface. This is done
by connecting Pin 7 (Remote) with ground (Pin 6) by
a jumper or switch. The status is then displayed like this:
Figure 15
Priorities
The analogue interface has priority over any other ope-
ration mode. Switching to external control can be done 
in  any  situation  (except  when  setup  menu  is  active). 
Hereby the set value inputs are activated and the set 
values for the load can only be adjusted by means of an 
external voltage source (PLC or 0...10V application) or 
with potentiometers. See table „8.4. Pin assignment of 
the analogue interface“) for an overview of the inputs. 
Additionally, the control via the interface card is blocked 
in this mode, but actual values can be read with it (i.e. 
monitoring).
8.1  Important notes
interface“.
Caution! The inputs are not protected against overvolta-
ges. Higher voltage than specified in section
 
input of the analogue interface may damage the device!
Please read and follow these instructions carefully:
•  Before connecting the hardware (with a Sub-D plug) 
which is used to control the analogue interface, wire 
all  necessary  connections  and  check  the  hardware 
for not supplying >10V respectively not more than the 
specified value.
•  The constant resistance regulation requires all four set 
values (U, I, P, R) to be fed in, if it‘s not used three set 
values (U, I und P) are sufficient.
•  By default, constant resistance regulation is activated! 
If not used, it should be deactivated by putting a bridge 
from pin 12 (R-active) to pin 6 (DGND):
Pin 12 = Low = Resistance regulation inactive 
Pin 12 = High = Resistance regulation active
•  If  resistance  regulation  is  used,  you  can  select  the 
resistance range before or while using the analogue 
interface. Pin 13 (R-Range) is used to switch between 
the two ranges:
Pin 13 = Low =Resistance range 2 is used 
Pin 13 = High = Resistance range 1 is used (default)
•  The input Rem-SB (Remote Standby, Pin 8) overrides 
the  pushbutton 
Input  on/off(4).  It  means,  that  you 
can switch the load input off with this pin at any time 
(even if the load was not set to external control via the 
analogue interface) and as long as this pin is tied to 0V 
(ground) the load input will be permanently off and can 
not be set to on with the pushbutton
Input on/off(4) or 
via remote control by an interface card.
•  The output VREF can be used to generate set values 
for the set value inputs VSEL, CSEL, PSEL and RSEL. 
For example: in case that only CC regulation is required, 
the set value input VSEL has to be tied to 0V, PSEL 
to VREF and CSEL can either be fed from an external 
0...10V source or via a potentiometer (GND and VREF, 
slider to CSEL). See also the table below.
•  Adjustable rise/fall times and pulse widths like in the 
Level A/B mode are not effective here. If a certain form 
of amplitude-time-progression is favoured, it has to be 
generated by an external function generator and fed in.
•  The  trigger  input  (Trigger  In)  has  no  function  when 
controlling the load via analogue interface (
External
mode
). That means that set value leaps have to be 
generated with the signal that is fed in to the set value 
inputs.
8.2  Example configurations
The table below shows example configurations for va-
rious  single  or  combined  regulation  modes. 
It always
applies, that pin 7 (Remote) always has to be pulled
to 0V (DGnd) and pin 12 (R-Active) too, if resistance
regulation is not used.
Explanation: it is not necessarily required to give a fixed 
10V to any non-variable input. It can of course also be 
a lower voltage in order to limit, for example, the power. 
Best to tie the non-variable inputs to VREF respectively 
to GND, if required.
Pin
VSEL
CSEL
PSEL
RSEL
R-active
Voltage regulation (CV)
var. 10V 10V -
L
Current regulation (CC)
0V var. 10V -
L
Power regulation (CP)
0V 10V var. -
L
Resistance regulation (CR) 0V 10V 10V var. H
Current & power (CC+CP)
0V var. var.  -
L
Current & resistance (CC+CR) 0V var. 10V var. H
„var.“ = variable set value of 0...10V, which can be, of course, pulsed 
in order to emulate the 
Level A/B operation.
„H/L“ = High or Low, for level see table „Pin assignment of the analogue 
interface“