Gw Instek PEL-3021 Electronic Load PEL-3021 데이터 시트
제품 코드
PEL-3021
OPERATION
83
Display
2.596
V
500.06
mA
1.2982
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
03/Sep/2012
V
V
UVP Alarm
Please Press Enter
To Clear Alarm
UVP
UVP indicator
Alarm message
when UVP is
tripped
when UVP is
tripped
Mode
CV
I Range
L 0.7A
V Range
L 15V
Response
Fast
Configure
OVP
Description
If the OVP is tripped, the PEL-3000 will turn off
the load.
the load.
The OVP levels can be set from 0V to 10%
higher than the rating voltage.
higher than the rating voltage.
Operation
1. Press
Main
> Configure[F5] > Protection[F1]
and set the OVP Level.
Range:
OVP Level: OFF, 0~Rating voltage +
10%
10%
Note: To turn OVP off, set the OVP voltage greater
than the current rating voltage + 10%.
than the current rating voltage + 10%.
Alarm
The OVP indicator and a message will only
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.
To clear the OVP indicator, remove the cause of
the over voltage ~ i.e., reduce the input voltage.
the over voltage ~ i.e., reduce the input voltage.