1.9.6.4  Resolution of the displayed values
In the display, set values can be adjusted in variable increments. The number of decimal places depends on the 
device model. The values have 3 to 5 digits. Actual and set values always have the same number of digits.
Adjustment resolution and number of digits of set values in the display:
Voltage,
OVP, UVD, OVD, 
U-min, U-max
Current,
OCP, UCD, OCD,
I-min, I-max
Power,
OPP, OPD,
P-max
Resistance,
R-max
Nominal
Digits
Min. 
Incre-
ment
Nominal
Digits
Min. 
Incre-
ment
Nominal
Digits
Min. 
Incre-
ment
Nominal
Digits
Min. 
Incre-
ment
80 V
4 0.01 V
22 A / 30 A
4 0.01 A
3.5 kW
3 10 W
4 Ω / 6 Ω
4 0.001 Ω
250 V
5 0.01 V
44 A / 60 A
4 0.01 A
7 kW
3 10 W
12 Ω
5 0.001 Ω
500 V
4 0.1 V
66 A / 70 A
4 0.01 A
10.5 kW
4 10 W
40 Ω / 60 Ω
4 0.01 Ω
750 V
4 0.1 V
90 A 
4 0.01 A
120 Ω / 160 Ω 5 0.01 Ω
1000 V
5 0.1 V
140 A / 170 A 4 0.1 A
240 Ω
5 0.01 Ω
1500 V
5 0.1 V
210 A 
4 0.1 A
360 Ω / 480 Ω 5 0.01 Ω
340 A / 510 A 4 0.1 A
550 Ω / 950 Ω 5 0.01 Ω
1100 Ω
4 1 Ω
1450 Ω
4 1 Ω
In manual operation every set value can be set in the increments given above. In this case 
the actual input values set by the device will lie within percentage tolerances as shown in the 
technical data sheets. These will influence the actual values. 
1.9.6.5  USB-Port (Front side)
The frontal USB port, located to the right of the rotary knobs, is intended for the connection of standard USB flash 
drives (flash drive). This can be used for:
Up or down loading from sequences for the arbitrary and UI-IU function generator
Updating HMI firmware (new languages, functions)
USB flash drives must be 
FAT32 formatted and have a maximum capacity of 32GB. All supported files must 
be held in a designated folder in the root path of the USB drive in order to be found. This folder must be named 
HMI_FILES, such that a PC would recognise the path G:\HMI_FILES if the drive were to be assigned the letter G. 
The control panel of the electronic load can read the following file types from a flash drive:
*.bin
Firmware updates only for the control panel. The file name format is given like 
96230058_FW-BE1_V114.bin, listed in short form as FW-BE1_V114.bin. Other 
files won’t be recognised or listed.
wave_u<arbitrary_text>.csv
wave_i<arbitrary_text>.csv
Function generator for an arbitrary function on voltage (U) or current (I)
The name must begin with wave_u / wave_i, the rest is user defined.
iu<arbitrary_text>.csv
IU table for the XY function generator.
The name must begin with iu, the rest can be user defined.
ui<arbitrary_text>.csv
UI table for the XY function generator.
The name must begin with ui, the rest can be user defined.