Maintaining the Test Tool 
 
Troubleshooting 
7
 
  
107 
Troubleshooting 
The Test Tool Shuts Down After a Short Time 
• 
The batteries may be empty. Check the battery 
symbol at the top right of the screen. A
symbol 
indicates that the batteries are empty and must be 
charged. Connect the BC190 power adapter. 
• 
The test tool is still on but the ‘display auto off’ timer is 
active, see Chapter 6 ‘Setting the Display AUTO-off 
Timer’. To turn the display on press any key (restarts 
the ‘display AUTO-off’ timer), or connect the BC190 
power adapter. 
• 
The power down timer is active, see Chapter 6 
‘Setting the Power Down timer’.  
Press 
 to turn the test tool on. 
The Screen Remains Black 
• 
Make sure that the test tool is on (press 
• 
You might have a problem with the screen contrast. 
Press 
USER
, then press 
F4
 . Now you can use 
the arrow keys to adjust the contrast. 
• 
The display ‘auto off’ timer is active, see Chapter 6 
‘Setting the Display AUTO-off Timer’. To turn the 
display on press any key (restarts the ‘display AUTO -
off’ timer), or connect the BC190 power adapter. 
The Test Tool Cannot Be Turned Off 
If the test tool cannot be turned off due to a software 
hangup, do the following: 
• 
Keep the ON/OFF key pressed for at least 5 seconds. 
FlukeView Does Not Recognize The Test Tool 
• 
Make sure that the test tool is turned on. 
• 
Make sure that the interface cable is properly 
connected between the test tool and the PC. Use only 
the test tool mini USB port for communication with a 
computer! 
• 
Make sure that no SAVE/RECAL/COPY/MOVE action  
from or to a USB-stick is being performed.   
• 
Make sure that the USB driver has been correctly 
installed, see Appendix A.