• On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
• When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks.
8. error messages / troubleshooting
In case of faults, the
message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement (in addition to
, the cardiac arrhythmia icon
appears in the
display),
4. the cuff tube is not inserted properly,
5. the cuff is inflated for longer than 16 seconds.
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.
9. Specifications
Model No.
BM 19
Method of Measurement Oscillometric, non-invasive blood
pressure measurement on the upper
arm
Range of Measurement Pressure 0 – 300 mmHg,
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
pulse 40 –199 beats/minute
Accuracy of the
systolic ± 3 mmHg /
pressure reading
diastolic ± 3 mmHg /
pulse ± 5 % of the displayed value
Measuring uncertainty
max. permissible standard deviation
according to clinical testing:
systolic: 8 mmHg /diastolic: 8 mmHg
Sets of Memory
2 x 60 memory places
Unit Dimension
L147 x W104 x H66 mm
Unit Weight
Approx. 310 g
Cuff Size
Wrist circumference approx.
23 – 33 cm
Operating Temperature + 10 °C to + 40 °C, ≤ 85 % RH
Storage-and
Storage-and
- 20 °C to + 70 °C, ≤ 85 % RH
transportemperature
Power Supply
Power Supply
4 x 1,5 V AA batteries (alkaline type
LR6)
Battery Life
For approx. 250 measurements,
depending on blood pressure level
depending on blood pressure level
or inflation pressure
Accessories
With storage pouch, instruction
manu al, 4 pcs “AA” battery
Classification
Application part Type BF
Protection class
internal power supply, IPXO,
no AP or APG, continuous operation
Key to symbols
Important! Read the operating
instruction!
Application part Type BF