• On no account must you immerse the computer in water, 
otherwise liquid can enter it and cause demage.
• When storing the device, make sure that no heavy objects 
are placed on top of it. Remove the batteries. The cuff tube 
should not have any sharp kinks.
8. error messages   / troubleshooting
In case of faults, the 
 message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement (in addition to 
, the cardiac arrhythmia icon 
 appears in the 
display),
4. the cuff tube is not inserted properly,
5. the cuff is inflated for longer than 16 seconds.
In the above cases, you must repeat the measurement. Make 
sure that the cuff tube is properly inserted and that you do 
not move or talk. Re-insert the batteries if necessary, or else 
replace them.
9. Specifications
Model No. 
BM 19
Method of Measurement   Oscillometric, non-invasive blood 
pressure measurement on the upper 
arm
Range of Measurement   Pressure 0   –   300 mmHg,  
systolic 40   –   280 mmHg,  
diastolic 40   –   280 mmHg,  
pulse 40   –199 beats/minute
Accuracy of the  
systolic ± 3 mmHg   /  
pressure reading  
 diastolic ± 3 mmHg   /  
pulse ± 5 % of the  displayed value
Measuring uncertainty 
 max. permissible standard deviation 
according to clinical testing:  
systolic: 8 mmHg /diastolic: 8 mmHg
Sets of Memory 
2 x 60 memory places
Unit Dimension 
L147 x W104 x H66 mm
Unit Weight 
Approx. 310 g
Cuff Size 
 Wrist circumference approx.  
23   –   33 cm
Operating Temperature   + 10 °C to + 40 °C, ≤ 85 % RH
Storage-and  
- 20 °C to + 70 °C, ≤ 85 % RH
transportemperature
Power Supply 
 4 x 1,5 V AA batteries (alkaline type 
LR6)
Battery Life 
 For approx. 250 measurements, 
 depending on blood pressure level 
or inflation pressure
Accessories 
 With storage pouch, instruction 
manu  al, 4 pcs “AA” battery 
Classification  
Application part Type BF
Protection class 
internal power supply, IPXO,
 
no AP or APG, continuous operation
Key to symbols 
 Important! Read the operating 
instruction! 
 
Application part Type BF